DMS-40PC
4½ D I G I T, L E D D I S P L AY D I G I TA L PA N E L V O LT M E T E R S
Performance/Functional Specifications
Typical at TA = +25°C and supply voltage = +5V using the single-ended input circuit,
unless otherwise noted.
Analog Inputs
Full Scale Input Range:
DMS-40PC-1
DMS-40PC-2
DMS-40PC-3
Input Impedance:
DMS-40PC-1
DMS-40PC-2, -3
Overvoltage Protection
Common Mode Voltage Range
CMRR (dc to 60Hz)
Control Inputs
Decimal Point Placement (Pins 4-6, 8)
Display Test (Pin 2)
Display Hold (Pin 9)
BCD Outputs
Logic levels (1 LSTTL load max.):
Logic "1"
Logic "0"
Performance
Sampling Rate
Accuracy (15 minute warm-up):
DMS-40PC-1 (VIN = +1.9V)
DMS-40PC-2 (VIN = +19V)
DMS-40PC-3 (VIN = +190V)
Zero Reading (VIN = 0 Volts)
Temperature Drift (0 to +50°C):
DMS-40PC-1
DMS-40PC-2, -3
Power Supply Requirements
Supply Voltage
Supply Current:
Standard Models
Low-Power Models
Display
Display Type and Size
Polarity Indication
Overrange Indication
Physical/Environmental
Operating Temperature
Storage Temperature
Humidity (Non-condensing)
Case Material
Weight
Min. Typ. Max. Units
--
±2
--
Volts
--
±20
--
Volts
--
±200
--
Volts
100
1000
--
M
0.8
1
--
M
--
--
±250 Volts
--
--
±2
Volts
--
86
--
dB
Tie to pin 3 to activate
Tie to pin 3 to activate all segments
Tie to pin 3 to hold last reading
+2.4
--
--
Volts
--
+0.4 +0.8 Volts
2.5 samples per second
--
--
--
"– 0001"
±2
±3
±3
"0000"
±3
±4
±4
"0001"
Counts
Counts
Counts
--
±0.4
±1 Cnts/°C
--
±0.4 ±1.5 Cnts/°C
+4.75 +5.00 +5.25 Volts
--
+100 +140
mA
--
+35
+50
mA
4½ Digit LED, 0.52"/13.2mm high
Autopolarity ("–" for negative VIN)
"– 0000" (flashing) for negative VIN
"0000" (flashing) for positive VIN
0
--
+50
°C
–20
--
+75
°C
0
--
95
%
Polycarbonate
0.75 ounces (21 grams)
Applies for transient or continuous overvoltages applied to (+) INPUT HI (pin 11)
with (–) INPUT LO (pin 12) properly connected. Pin 12 is not overvoltage protected
(see Figure 1). Voltages applied to pin 12 should not exceed the supply voltage.
See Technical Notes.
BCD outputs are optional and must be specified in the part number.
See Ordering Information.
Includes high-intensity and BCD-output models.
Order on-line at www.datel.com
Ordering Information
DMS-40PC - 1 - R S
Input Range:
1 = ±2V
2 = ±20V
3 = ±200V
LED Color:
GS = Standard Green
RH = High-Intensity Red
RL = Low-Power Red
RS = Standard Red
YS = Standard Yellow
BCD Output Models:
BCD outputs are only available on standard red meters.
DMS-40PC-1-RS-BCD for ±2V input range
DMS-40PC-2-RS-BCD for ±20V input range
DMS-40PC-3-RS-BCD for ±200V input range
DMS-40PC-1-RL-BCD for ±2V input range
DMS-40PC-2-RL-BCD for ±20V input range
DMS-40PC-3-RL-BCD for ±200V input range
Accessories:
DMS-PS1-CM
DMS-30-CP
DMS-BZL1
DMS-BZL2
DMS-EB
+5V/1.0A AC/DC power supply module
Panel cutout punch
DMS-40 bezel assembly
DMS-40 bezel assembly with sealing gasket
Application/evaluation board with standard
MOLEX connector, decimal point solder pads
and attenuation resistor pads.
A panel-mount retaining clip is supplied with each model.
Technical Notes
1. ANALOG COMMON (Pin 10): This pin is an internal, low-noise
ground for the DMS-40PC. It is internally connected to pin 3
(5V RETURN). Do not connect pin 10 to either pin 3 or your
system ground as this will create a ground loop and possibly result
in erroneous readings.
2. REFERENCE INPUT/OUTPUT (Pin 7): This pin accesses the
meter's internal reference and is used during the factory calibration
procedure. Pin 7 should be left open in most common applications.
It can be used in certain "ratiometric" applications in which it is
desirable for the meter’s reference to track an external reference.
See Ap Note 3 in the DATEL Panel Meter Catalog for more details.
3. DISPLAY TEST (Pin 2): Connecting pin 2 to ground (pin 3, 5V
RETURN) will activate all LED segments, and the display will read
"–18888" regardless of the actual applied input. To reduce self-
heating, the display should not be left in the "test" mode for
more than 10 seconds. This pin should be left open if unused.
4. DISPLAY HOLD (Pin 9): For normal operation, this pin should be
left open. To hold the meter’s last reading and display it continu-
ously, tie pin 9 to ground (pin 3, 5V RETURN).
5. Decimal Point Placement: The location of the decimal point is
user-selectable, and the decimal point control pins (DP1-DP4) are
active low functions. Select the appropriate decimal point by tying
pin 4, 5, 6 or 8 to pin 3 (5V RETURN). Unused decimal point
location pins should be left open.