DatasheetQ Logo
Electronic component search and free download site. Transistors,MosFET ,Diode,Integrated circuits

B1541A-001 データシートの表示(PDF) - HP => Agilent Technologies

部品番号
コンポーネント説明
メーカー
B1541A-001
HP
HP => Agilent Technologies HP
B1541A-001 Datasheet PDF : 23 Pages
1 2 3 4 5 6 7 8 9 10 Next Last
Voltage range, resolution, and accuracy (high speed ADC)
Voltage range Force resolution Measure resolution Force accuracy¹
Measure accuracy¹
±0.5 V
25 µV
0.5 µV
±(0.018 % + 150 µV) ±(0.01 % + 250 µV)
±2 V
100 µV
2 µV
±(0.018 % + 400 µV) ±(0.01 % + 700 µV)
±5 V
250 µV
5 µV
±(0.018 % + 750 µV) ±(0.01 % + 2 mV)
±20 V
1 mV
20 µV
±(0.018 % + 3 mV)
±(0.01 % + 4 mV)
±40 V
2 mV
40 µV
±(0.018 % + 6 mV)
±(0.015 % + 8 mV)
±100 V
5 mV
100 µV
±(0.018 % + 15 mV) ±(0.02 % + 20 mV)
1. ± (% of read value + offset voltage V)
2. 100 mA (Vo ≤ 20 V), 50 mA (20 V < Vo ≤ 40 V), 20 mA (40 V < Vo ≤ 100 V), Vo is the output voltage in Volts.
Maximum current
100 mA
100 mA
100 mA
100 mA
²
²
Current range, resolution, and accuracy (high speed ADC)
SMU type
Current Force
Measure
range resolution resolution¹,²
Force accuracy³
Measure accuracy³
Maximum
voltage
HRSMU w/ ASU ±1 pA
1 fA
100 aA
±(0.9 %+15 fA)
±(1.8 %+12 fA)
100 V
HRSMU
±10 pA
5 fA 400 aA (with ASU) ±(0.46 %+30 fA+10 aA x Vo) ±(0.5 %+15 fA+10 aA x Vo)
100 V
1 fA (HRSMU)
±100 pA
5 fA 500 aA (with ASU) ±(0.3 %+100 fA+100 aA x Vo) ±(0.5 %+40 fA+100 aA x Vo) 100 V
2 fA (HRSMU)
MPSMU ±1 nA 50 fA
10 fA
±(0.1 %+300 fA+1 fA x Vo) ±(0.25 %+300 fA+1 fA x Vo) 100 V
±10 nA 500 fA
10 fA
±(0.1 %+3 pA+10 fA x Vo) ±(0.25 %+2 pA+10 fA x Vo)
100 V
±100 nA
5 pA
100 fA
±(0.05 %+30 pA+100 fA x Vo) ±(0.1 %+20 pA+100 fA x Vo) 100 V
±1 µA 50 pA
1 pA
±(0.05 %+300 pA+1 pA x Vo) ±(0.1 %+200 pA+1 pA x Vo) 100 V
±10 µA 500 pA
10 pA
±(0.05 %+3 nA+10 pA x Vo) ±(0.05 %+2 nA+10 pA x Vo) 100 V
±100 µA
5 nA
100 pA
±(0.035 %+15 nA+100 pA x Vo) ±(0.05 %+20 nA+100 pA x Vo) 100 V
±1 mA 50 nA
1 nA
±(0.04 %+150 nA+1 nA x Vo) ±(0.04 %+200 nA+1 nA x Vo) 100 V
±10 mA 500 nA
10 nA
±(0.04 %+1.5 µA+10 nA x Vo) ±(0.04 %+2 µA+10 nA x Vo) 100 V
±100 mA
5 µA
100 nA
±(0.045 %+15 µA+100 nA x Vo) ±(0.1 %+20 µA+100 nA x Vo)
4
1. Specified measurement resolution is limited by fundamental noise limits. Minimum displayed resolution is 1 aA at 1 pA range by 6 digits.
2. Measurements made in the lower ranges can be greatly impacted by vibrations and shocks. These specifications assume an environment free of
these factors.
3. ± (% of read value + offset current (fixed part determined by the output/measurement range + proportional part that is multiplied by Vo))
4. 100 V (Io ≤ 20 mA), 40 V (20 mA < Io ≤ 50 mA), 20 V (50 mA < Io ≤ 100 mA), Io is the output current in Amps.
Power consumption
Voltage source mode
Voltage range
Power
0.5 V
20 x Ic (W)
2V
20 x Ic (W)
5V
20 x Ic (W)
20 V
20 x Ic (W)
40 V
40 x Ic (W)
100 V
100 x Ic (W)
Where Ic is the current compliance setting.
Current source mode
Voltage compliance Power
Vc ≤ 20
20 x Io (W)
20 < Vc ≤ 40
40 x Io (W)
40 < Vc ≤ 100
100 x Io (W)
Where Vc is the voltage compliance setting
and Io is output current.
MPSMU and HRSMU measurement and output range
Current (mA)
100
50
20
-100
-40 -20
20 40
-20
-50
-100
100
Voltage (V)
4