DatasheetQ Logo
Electronic component search and free download site. Transistors,MosFET ,Diode,Integrated circuits

MVS5410 データシートの表示(PDF) - PerkinElmer Inc

部品番号
コンポーネント説明
メーカー
MVS5410
PerkinElmer
PerkinElmer Inc PerkinElmer
MVS5410 Datasheet PDF : 2 Pages
1 2
MVS-5410 Strobe
Optical Specifications
Spectral bandwidth
Flashlamp life (5)
Flash duration (1)
Intensity level
Illumination Characteristics
Distance
1 foot
2 feet
3 feet
Area Illuminated
15” x 15”
30” x 30”
45” x 45”
MVS-5410-10
300 to 1100+ nm
>108 flashes
20 microseconds
Adjustable from 35% to 100%
Photometric
38 lux-sec
10 lux-sec
4 lux-sec
Radiometric
32 µJ/cm2
9 µJ/cm2
4 µJ/cm2
MVS-5410-02
300 to 1100+ nm
>108 flashes
20 microseconds
Adjustable from 35% to 100%
Area Illuminated
15” x 15”
30” x 30”
45” x 45”
Photometric
13 lux-sec
3.5 lux-sec
1.5 lux-sec
Radiometric
12 µJ/cm2
3 µJ/cm2
1.5 µJ/cm2
Note 1:
Note 2:
Note 3:
Note 4:
Note 5:
Energy per flash is constant up to 100Hz; total output power is constant from 100Hz to 200Hz. (Power = Energy x Flashrate)
Energy per flash is constant up to 200Hz; total output power is constant from 200Hz to 400Hz. (Power = Energy x Flashrate)
At maximum intensity. Flash to Flash Variation ±5%
Lux = Lumen-second/meter2
Where light output is not less than 50% of maximum value.
Electrical Specifications
Input voltage
Input current
Power to Lamp
External trigger:
Pulse
Current
Duty cycle
Flash delay
Internal trigger
90-230 VAC 50/60 Hz
135 W rms maximum
100 W rma maximum
+5 volt TTL pulse, 10-100
µsec into an opto-isolator
with internal 150 ohm
nominal resistor
20 mA
up to 90%
18µsec max from leading
edge of trigger pulse
Variable via 10-turn
potentiometer
Environmental Specifications
Operating temperature
Storage temperature
Shock and vibration
-10 to +110°F (-23 to +43°C)
-40 to +194°F (-40 to +90°C)
1.5G, 50 to 200 Hz per
MIL-STD-810C
Controls
Flash rate intensity (1 turn potentiometer) internal/external
Flash rate (10 turn potentiometer) internal only
Internal/External trigger switch
Flash standby (stopps flashlamps operation) internal/external
AC Power On/OFF switch (back side of unit)
Indicators
Trigger light
AC power light
Mechanical Specifications
10.75 (273)
0.281 DIA. HOLE, 2 PLACES
FOR MOUNTING
7.50 (191)
Intensity
RATE
Trigger
INT
EXT
POWER
O
STBV
10.35 (263)
6.78
(172)
6.30
(160)
12.25 (311)
9.71
(247)
ooooooo
oooooooooooooooo
ooooooo
oooooooooooooooo
ooooooo
oooooooooooooooo
ooooooo
oooooooooooooooo
Air Exhaust
AC
Input
Cooling Fan
On/Off
Switch
Control
Input
Rear Panel
DIMENSIONS ARE IN
INCHES (MILLIMETERS)
WEIGHT: 7.2 LBS (3.27kg)
CAUTION
Some glass flashlamps are under
high internal pressure, and, if bro-
ken, could result in glass particles
being blown into the face and hand
areas. To prevent injury, wear suit-
able protective devices such as
safety glasses and/or face mask
and gloves.
Some types of pulsed lamps gen-
erate intense ultraviolet radiation
which, if not properly shielded from
personnel in the area, will cause
burns to any exposed skin and
especially to the eyes. Do not
expose any skin area or the eyes
to the direct or reflected radiation
of an operating lamp. If you have
to view an operating lamp, always
use protective covering for
exposed skin area and ultraviolet-
attenuating goggles for the eyes.
For more information e-mail us at opto@perkinelmer.com or visit our web site at www.perkinelmer.com/opto.
All values are nominal; specifications subject to change without notice.
USA:
Europe:
Asia:
PerkinElmer Optoelectronics
PerkinElmer Optoelectronics
PerkinElmer Optoelectronics
35 Congress Street
Wenzel-Jaksch-Str. 31, D-65199 47 Ayer Rajah Crescent #06-12
Salem, MA 01970
Weisbaden, Germany
Singapore 139947
Toll Free: (800) 950-3441 (USA)
Phone: +49 (0) 611-492-437
Phone: +65 775 2022
Phone: (978) 745-3200
Fax: +49 (0)611-492-159
Fax: +65 775 1008
.
Fax: (978) 745-0894
© 2001 PerkinElmer inc. All rights reserved.
0601
www.perkinelmer.com/opto

Share Link: 